IVDR Compliance: Key Challenges and Opportunities for IVD Manufacturers in Europe
Article Summary
The key IVDR challenges facing IVD manufacturers in Europe, from defining intended purpose and generating clinical evidence to preparing technical documentation and engaging with notified bodies.Article Contents
The Changing Role of IVD Manufacturers
The transition to the In Vitro Diagnostic Regulation (IVDR) has raised the bar for manufacturers seeking access to the European market. From defining intended purpose and generating clinical evidence to navigating notified body assessments, the head of the IVD Notified Body from BSI explains where manufacturers are getting things wrong and what they can do differently.
For in vitro diagnostic manufacturers, getting a product to market in Europe requires more than demonstrating that a test works. Manufacturers must understand how their devices fit within the regulatory framework, establish the evidence needed to support their claims and demonstrate that they meet the requirements of the In Vitro Diagnostic Regulation.Â
For some companies, particularly those with limited past exposure to medical device regulation, this represents a significant learning curve.Â
Speaking at the Veeva MedTech Summit in Amsterdam, Alex Laan, Head of the IVD Notified Body at BSI, identified one of the biggest challenges facing manufacturers: understanding the European regulatory system and translating its requirements into practical steps.
Some manufacturers continue to view their products primarily as laboratory tests. However, IVDs generate information that informs clinical decisions, influences treatment pathways and ultimately affects patient outcomes. Recognising their role as medical devices, rather than simply laboratory products, is fundamental to understanding the regulatory responsibilities that follow.Â
As regulatory expectations increase, manufacturers need to understand not only what is required, but how to demonstrate compliance.

1. Understanding IVDR Requirements and Regulatory Expectations
The introduction of the IVDR has fundamentally changed the regulatory setting for IVD manufacturers.Â
Under the previous In Vitro Diagnostic Medical Devices Directive (IVDD), most manufacturers could self-declare their devices before placing them on the market. The IVDR has brought a much greater proportion of IVDs within the scope of notified body conformity assessment.Â
For established manufacturers, this represents a considerable change. Some question why products that have been on the market for many years must now undergo more extensive assessment and provide additional clinical evidence.Â
The BSI expert acknowledges that this is an understandable concern from a manufacturer’s perspective. However, a history of successful use does not remove the responsibility to demonstrate that a device meets the applicable regulatory requirements.Â
The challenge is particularly significant for manufacturers who have had limited exposure to medical device regulation. Uncertainty about how to interpret the requirements, what evidence is needed and where to seek clarification can make the process appear more complicated than it needs to be.Â
There can also be reluctance to approach notified bodies with questions. Manufacturers may be concerned about sharing information or believe that raising particular challenges could have negative consequences.Â
Alex Laan encourages greater openness, pointing out that manufacturers frequently encounter similar challenges.Â
Sharing experiences and discussing uncertainties can help companies recognise that their difficulties are not necessarily unique. Notified bodies can also provide direction, explain regulatory expectations and help address assumptions, even though they cannot tell manufacturers how to achieve compliance.Â
2. Defining Intended Purpose: Why It Matters for Clinical Evidence
One of the areas where IVD manufacturers most frequently encounter difficulties is clinical evidence. A common underlying issue is the definition of intended purpose.Â
Before placing a device on the market, manufacturers need to establish what it is intended to do and demonstrate that it can meet the performance requirements associated with that purpose.Â
Intended purpose is much more than a description in the technical documentation. It influences the device’s performance requirements, the clinical evidence strategy and the claims a manufacturer can make.Â
The consequences become clearer when considering a diagnostic assay used for different clinical purposes.Â
Take a syphilis assay, for example. The same underlying technology could be used for screening or confirmatory testing, but the performance expectations would differ.Â
For a screening assay used to identify syphilis antibodies in a blood donation, sensitivity is particularly important. The test needs to minimise the risk of false-negative results, as failing to identify a positive result could have serious consequences for the recipient.Â
A confirmatory assay serves a different purpose. It is used in a different clinical context, where there is already an indication that a person may have the disease. Here, specificity becomes particularly important, and the performance requirements may differ from those of a screening test.Â
The technology alone does not determine what a device must demonstrate. Its intended purpose, the setting in which it will be used and the expectations of its users all influence the required performance.Â
This is why manufacturers need to define their intended purpose carefully and ensure that their clinical evidence supports it. An unclear or poorly formulated intended purpose can undermine the entire evidence strategy.

3. Generating Clinical Evidence for Innovative IVDs
Demonstrating performance can be particularly challenging for innovative diagnostic technologies.Â
For any IVD placed on the market, manufacturers need to establish its scientific validity and determine its analytical and clinical performance. This requires appropriate evidence, testing and technical documentation to support the device’s intended purpose.Â
Established products may have an advantage in this respect. Manufacturers may already have confirmation that the analyte is accepted and that the test works in practice, including recognition by diagnostic laboratories.Â
Additional clinical evidence and testing may still be necessary for these devices, but there is an existing foundation on which to build.Â
With genuinely novel technologies, there are more unanswered questions. Manufacturers must establish the scientific and clinical basis for the test, making the evidence-generation process more demanding.Â
The challenge is therefore twofold: developing the necessary data and understanding how to demonstrate compliance with the regulatory requirements.Â
For manufacturers unfamiliar with the regulatory framework, these demands can create further uncertainty and delay.
4. IVDR Vigilance: Understanding the Impact of Diagnostic Errors
Incident assessment and reporting is another area that can present difficulties for IVD manufacturers.Â
Unlike many other medical devices, where a device failure may have a relatively direct relationship with an adverse patient outcome, the consequences of an IVD error can be less straightforward.Â
For example, if an implant fails or causes a complication, there may be a clear connection between the device and the patient’s condition.Â
With an IVD, the impact is often indirect. An incorrect result may be interpreted alongside other clinical information, and a patient may undergo further testing before the implications become apparent.Â
This can make it more difficult to establish whether an incident has occurred, understand its relationship with the device and collect the information needed to assess it appropriately.Â
Nevertheless, the indirect nature of the impact does not make vigilance any less important. IVD results inform clinical decisions, and manufacturers need to understand how incorrect results could affect patient management.Â
Developing appropriate processes for reviewing incidents and identifying potential relationships between device performance and patient outcomes is therefore essential.
5. Why Early Notified Body Engagement Matters
When should manufacturers involve a notified body in their regulatory planning?Â
Alex Laan’s answer is straightforward: as early as possible.Â
Early engagement gives manufacturers an opportunity to test their assumptions, understand regulatory expectations and identify potential gaps before they become more difficult to address.Â
It can also make commercial sense. Europe represents a significant market opportunity, and CE certification can support access to other markets that recognise or leverage European conformity assessment.Â
For manufacturers planning to enter Europe, initial conversations with a notified body can help establish a clearer regulatory direction.Â
BSI offers different forms of structured dialogue, including discussions tailored to individual manufacturers. These provide opportunities to explore regulatory questions and clarify expectations ahead of formal conformity assessment.Â
Timing is particularly important for manufacturers still relying on the provisions, progressing into the IVDR.Â
Although the transition arrangements have given eligible manufacturers additional time, Alex Laan warns against treating this as an opportunity to postpone preparation indefinitely.Â
Some manufacturers may be tempted to make full use of the available transition period before beginning their IVDR conformity assessment. However, as the applicable deadline approaches, the number of manufacturers seeking assessment could create additional pressure.Â
If the necessary conformity assessment is not completed in time, manufacturers may lose the ability to place affected legacy devices on the market.Â
The consequences could extend beyond the manufacturer, potentially affecting laboratories, hospitals and the availability of diagnostic tests.Â
The message is clear: transitional provisions provide time to prepare, not a reason to delay.Â

6. What IVD Manufacturers Should Do Differently
With discussions about potential changes to the IVDR continuing, manufacturers may be tempted to wait and see what the future regulatory framework will look like before committing to their conformity assessment strategy.Â
The BSI expert advises against this approach.Â
Although proposals to amend elements of the regulation are being discussed, manufacturers should not assume that the fundamental requirements will change dramatically.Â
His recommendation is to start preparing for the IVDR now rather than waiting for a possible future version of the regulation.Â
Alex Laan also offers three practical recommendations for manufacturers preparing their technical documentation and submissions.Â
Make sure the submission is compliant.Â
Manufacturers need to understand the applicable regulatory requirements and ensure their technical documentation addresses them. A clear understanding of the regulatory roadmap helps establish what evidence is needed and where potential gaps may exist.Â
Provide context, not just documentation.Â
A technical file should do more than present information. Manufacturers should explain the product’s history, relevant design iterations, the evidence they have relied upon and any additional studies undertaken to address identified gaps.Â
For example, if a manufacturer has relied on legacy clinical or performance data but identified areas requiring further evidence, explaining how those gaps were addressed gives the notified body a clearer understanding of the submission.Â
Similarly, if certain requirements or elements are not applicable to a particular device, manufacturers should explain why.Â
Be open and transparent.Â
A well-structured submission should help the reviewer understand the manufacturer’s reasoning and follow the evidence without repeatedly having to ask where information can be found or why particular decisions were made.Â
Providing context and being transparent about the development and evidence-generation process allows notified body technical reviewers to concentrate on the substantive aspects of the assessment.Â
Making IVDR Conformity Assessment More Efficient
A more interactive approach to technical documentation review can also help manufacturers prepare for conformity assessment.Â
BSI has developed an interactive dedicated service that allows manufacturers to sit down with its team and review their technical files together. The aim is to identify potential gaps, clarify information and address questions at the start of the review process.Â
By resolving straightforward queries earlier, the process can become more focused, allowing reviewers to concentrate on the issues that matter and helping to shorten the overall review and certification process.Â
Ultimately, successful IVDR compliance requires manufacturers to take a proactive approach. A clear intended purpose, appropriate clinical evidence, well-structured technical documentation and early notified body engagement all contribute to a more manageable conformity assessment process.Â
For manufacturers planning to access or maintain their presence in the European market, the time to prepare is now.
Disclaimer. The views and opinions expressed in this article are solely those of the author and do not necessarily reflect the official policy or position of Test Labs Limited. The content provided is for informational purposes only and is not intended to constitute legal or professional advice. Test Labs assumes no responsibility for any errors or omissions in the content of this article, nor for any actions taken in reliance thereon.