Medical Device Disinfection Validation
Frequently Asked Questions
Do we need to validate disinfection if we’ve already validated cleaning?
Yes. Cleaning and disinfection are separate, complementary steps: cleaning must be validated first, and disinfection must then be validated to show the process delivers the intended microbial risk reduction for your device and use conditions.
What’s the difference between validating manual vs automated disinfection?
Manual disinfection: we verify the IFU’s contact time, temperature, chemistry, and method are effective and reproducible under worst‑case, operator‑to‑operator variation.
Automated disinfection (e.g., washer‑disinfectors): we follow the ISO 15883 series to demonstrate performance with defined loads, utilities, and parameters.
Which standards and guidance do you align to for disinfection?
Core references are: ISO 17664‑1/‑2 (what manufacturers must provide in IFUs), the FDA 2015 Reprocessing Guidance (how to validate reprocessing, including disinfection), and the ISO 15883 series for automated cleaning/disinfection systems.
What does a disinfection validation actually look like?
We run simulated‑use testing (if applicable) that reflects your device design, typical soils, and worst‑case conditions; we then execute your IFU’s manual or automated disinfection steps and document that the process parameters and outcomes meet recognised requirements (FDA guidance + ISO 15883 for automated systems).
How does cleaning interact with disinfection in the study?
Effective disinfection depends on effective cleaning: residual soils can shield microbes and invalidate disinfection claims. That’s why regulators expect cleaning to be validated and documented before (or alongside) disinfection validation in the technical file.
We won’t keep you waiting
We respond to all enquiries within minutes of receiving them.
Our lead times match the urgency and demand in the market – we offer industry-leading turnaround times because we know how critical reliable data is for your project. While what you’ve seen is our typical lead time, the final timeline will depend on your specific study requirements. But rest assured, we’ll do everything we can to deliver even faster.
Think of it like a relay race – we’ll hand over the data baton as quickly as possible, so you can keep moving forward on your medical device journey.
Explore other relevant testing services
Medical Device Cleaning Validation
We validate your medical device cleaning instructions, as detailed in your IFU, which can be manual or automated.
- 4-5 week typical lead time
- Simulated use/accumulation study
- BS EN ISO 17664-1:2021
- ANSI/AAMI ST98/ISO 15883-5
- ASTM F3208-20
Medical Device Sterilisation Validation
We validate your medical device sterilisation instructions, which can be moist heat, low temperature hydrogen peroxide or other process.
- 4-5 week typical lead time
- Simulated use/accumulation study
- BS EN ISO 17664-1:2021
- ISO 14937:2009
- BS EN ISO 17665:2024
Medical Device Biocompatibility Testing
We offer a comprehensive range of biocompatibility tests to evaluate potential biological risks arising from your medical device.
- 4-5 week typical lead time
- Biological Evaluation Plan & Report
- In Vitro & In Vivo methods
- Extractables & Leachables (E&L) Studies
- Also testing following reprocessing procedures or accelerated aging to meet lifecycle requirements